PI V-308 Voice Coil PIFOC Objective Focus Drive

V-308 is a vertical objective-focus positioner with a voice-coil motor, crossed-roller guide and adjustable magnetic weight compensation. The current V-308.753030 specification provides 7 mm travel with user-adjustable travel limitation and compensation for a load up to 1 kg under the documented orientation. Objective holders and adapters must be specified separately.
Key Specifications
Travel | 7 mm; adjustable limitation to total 1–7 mm |
Drive / guide | Voice coil / crossed roller with cage control |
Weight compensation | User-adjustable magnetic compensation up to 1 kg in specified arrangement |
Sensor resolution | 1 nm incremental linear feedback |
Minimum incremental motion | 10 nm typical |
Settling example | 15 ms for 100 or 250 nm step in±15 nm error band under stated conditions |
Motor force | 4.2 N typical;9.45 N peak in documented direction scope |
Applications and Industries
Microscopy, wafer/surface inspection and precision objective focusing with a defined optical and mechanical load.
Electronics and semiconductors; Medical and life sciences
Review Summary
Consider V-308 for fast microscope focus or surface inspection when the objective mass and optical geometry fit its load and travel. Its 1 nm feedback resolution is separate from 10 nm minimum incremental motion and from travel-wide positioning accuracy. Small-step settling figures apply only to their stated step and error band.
Price and Configuration
Request a quote for the selected configuration,quantity and delivery schedule. Confirm the exact model,options,integration scope and support terms.